XP2972 PMT History

CBS - Cathode Blue sensitivity ABS - For a anode blue sensitivity Id - Dark current

SN CBS [uA/lmF] ABS [A/lmF] HV [V] Id[nA] Gain [106] Comment
34391 12.6 10 1188 4.98 0.79 new, will be installed in TE1
29181 used, low gain, was in S07DL1R

GS06DL3U

Reason for the service: The detector counts more than 10 times compare to the detectors above, left and right of it. Will examine it if:

  • The detector is light tight
  • If the AVD is damaged
  • If there is radiation damage to the scintillator (not related to the above problem)
  1. LED check with multimeter:
    • Forward bias 1.75 V
    • Backward bias, open
  2. Dark counts, trigger 30 mV :
HV [V] Ileakage [uA] Counts[Hz]
300 200 0
1000 672 0
1500 1008 30
  • Visual inspection of the SCI
    • No yellowing, signs of mechanical damage on the surface.
    • Under UV, no spots

The Standard operating voltage is 1100 V, Il = 739 uA, counting rate bellow 5 Hz