User Tools

Site Tools


instruments:n-fct:overview

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
instruments:n-fct:overview [2019/08/09 13:10]
reeg
instruments:n-fct:overview [2021/11/24 16:41] (current)
reeg
Line 1: Line 1:
 +**
 +Basic BEA papers about FCTs, and special info for FCTs in FAIR HEBT lines:**
  
-Already in 1995, four GSI-made FCTs T-DS114 had been installed in the GSI HEST lines+[[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=instruments:n-fct:fct_electronicschain.pdf|Calculated Charge Resolutions for Fast Current and Resonant Transformers]]
  
-**GTE1DT1FP, GTE5DT1FP, GHHTDT6FP, GHTPDT1FP**+[[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=instruments:n-fct:transformer_resolution.pdf|Signal Estimates and Electronics Scheme for Fast Current Transformers in High-Energy Transfer Lines]]
  
-After 2005, commercially available FCTs from [[https://www.bergoz.com/fct?bs=5|Bergoz Instrumentation]] have been added to SIS 18 and ESR: +[[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=instruments:n-fct:fct_pulseattenuation.pdf|Signal Distortion in Coaxial Transmission Lines: Application to Logic Pulses, Photomultiplier Pulses, and Fast Current Transformer Signals]]
  
-**GS01DT1FP, GE02DT1FP**+[[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=instruments:n-fct:fct-response--htp.pdf|Investigation of a little bit of 
 +Fast Current Transformer response at HTP]]
  
-Basic informations are given in this paper: [[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=ed:dipac2001_ps-08.pdf|DIPAC 2001]]+---- 
 +** 
 +Already in 1995, four GSI-made FCTs T-DS114 had been installed in the GSI HEST lines:**  
 + 
 +**GTE1DT1FP\\ GTE5DT1FP\\ GHHTDT7FP\\ GHTPDT1FP** 
 + 
 +After 2005, commercially available FCTs from [[https://www.bergoz.com/fct?bs=5|Bergoz Instrumentation]] have been added to SIS 18 and ESR:  
 + 
 +**GS01DT1FP**, which had been relocated from it's initial position in GS07, eventually**\\ GE02DT1FP**
  
 ---- ----
Line 14: Line 25:
 **Main characteristics of the installed FCTs:** **Main characteristics of the installed FCTs:**
  
-**GSI FCT HEST** +**GSI FCT in HEST lines** 
-  * model no.: T-DS114 +  * GSI model no.: T-DS114 
-  * ID = 160mm +  * ID of toroid = 160mm 
-  * sensitivity/ransimpedance: 0.23 V/A +  * sensitivity/transimpedance: 0.23 A/V (or 4.2 V/A) 
-  * termination: 50 Ω +  * @termination: 50 Ω 
-  * low freq. cutoff: ~2 kHz +  * low freq. cutoff -3dB: ~2 kHz 
-  * high freq. cutoff: ~300 MHz+  * high freq. cutoff -3dB: ~300 MHz
  
 [[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=ed:t-ds114.pdf|GSI FCT datasheet]] [[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=ed:t-ds114.pdf|GSI FCT datasheet]]
  
-**Bergoz FCT SIS18** +**Bergoz FCT in SIS18** 
-  * model no.: FCT-260-50:1-LD-H +  * Bergoz model no.: FCT-260-50:1-LD-H 
-  * ID = 260 mm+  * ID of toroid = 260 mm
   * sensitivity/transimpedance: 0.5 V/A   * sensitivity/transimpedance: 0.5 V/A
-  * termination: 50 Ω+  * @termination: 50 Ω   
 +  * low freq. cutoff -3dB: ~200 Hz 
 +  * high freq. cutoff -3dB: ~690 MHz
   * <color #ed1c24>separate output for Ring RF through power divider, which reduces sensitivity by factor 2 !</color>   * <color #ed1c24>separate output for Ring RF through power divider, which reduces sensitivity by factor 2 !</color>
  
 [[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=ed:sis-fct-260-50_report.pdf|documents sensor #1768]] [[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=ed:sis-fct-260-50_report.pdf|documents sensor #1768]]
 +
 +[[https://www-bd.gsi.de/dokuwiki/lib/exe/fetch.php?media=instruments:n-fct:wepf18.pdf|A FESA DAQ FOR FAST CURRENT TRANSFORMER IN SIS 18]]
  
 {{:instruments:n-fct:fct-toroid.png?nolink&400|}} {{:instruments:n-fct:fct-toroid.png?nolink&400|}}
Line 37: Line 52:
 Typical Bergoz FCT toroid Typical Bergoz FCT toroid
  
-**Bergoz FCT ESR** +**Bergoz FCT in ESR** 
-  * model no.: FCT-260-1.25-LD +  * Bergoz model no.: FCT-260-1.25-LD 
-  * ID = 260 mm+  * ID of toroid = 260 mm
   * sensitivity: 1.25 V/A   * sensitivity: 1.25 V/A
-  * termination: 50 Ω+  * @termination: 50 Ω 
 +  * low freq. cutoff -3dB: ~650 Hz 
 +  * high freq. cutoff -3dB: ~570 MHz
   * <color #ed1c24>separate output for Ring RF through power divider, which reduces sensitivity by factor 2 !</color>   * <color #ed1c24>separate output for Ring RF through power divider, which reduces sensitivity by factor 2 !</color>
  
instruments/n-fct/overview.1565349053.txt.gz · Last modified: 2019/08/09 13:10 by reeg