Line 8: Line 8:
  
 Bellow is a spectrum taken with the HDO6104A scope when for a BLM set to count 3 kHz (in this case XP2972 34391) Bellow is a spectrum taken with the HDO6104A scope when for a BLM set to count 3 kHz (in this case XP2972 34391)
 +
 +{{:hi:hi:sn34391-137cs-1275v--00000.jpg?700|}}
  
 ====== XP2972 PMT History ====== ====== XP2972 PMT History ======
Line 38: Line 40:
     * No yellowing, signs of mechanical damage on the surface.     * No yellowing, signs of mechanical damage on the surface.
     * Under UV, no spots     * Under UV, no spots
-The Standard operating voltage is 1100 V, I<sub>l</sub> = 739 uA, counting rate bellow 5 Hz +The Standard operating voltage is 1100 V, I<sub>l</sub> = 739 uA, counting rate bellow 5 Hz 
 + 
 + 
 +====== FAIR Type  ====== 
 +   * Measurement with Micsing scope, BW 150 MHz, dT=100 ns/div, dV = 200 mV/div, trigger -104 mV  
 +   * The LED drive, pulse, positive, voltage is 3V, width 50 ns, rise/fall time 5 ns, frequency 1 kHz 
 + 
 +^ PMT S/N ^ AVD S/N [uA] ^ I@1500V [uA] ^ Dark counts@1500V [Hz]^ Vmin[V] LED on ^ 
 +| LA8177  | 2019-101     | 540          |           < 15        |     1.29    | 
 +| LA8178  | 2019-100     | 570          |           < 15        |     1.35    | 
 +| LA8179  | 2019-102     | 540          |           < 10        |     1.36    | 