Capacitance Measurements

The goal is to measure the capacitance of the detectors which are in use in the high energy beam lines as a fiction of frequency. This information can be used in to simulate detector response and for diagnostics of the detector health. For instance defect cables can be discovered by controlling the detector capacitance.

Measuring equipment and measurement procedure

Keysight U1733C

Keysight N9917A

For S11 measurement, select NA and set Start F = 30 kHz, Stop F = 10 MHz

For all measurements perform quick calibration with SMA open and 50 Ohm load from the Nano-VNA set

Detector C for low frequencies

Detector F [kHz] C [pF] Θ [deg]
Ref.Cal. L=65cm 0.100 66.1 -89.0
0.120 67.2 -89.7
1 67.1 -89.8
10 67.0 -80.0
100 66.9 -89.1
FAIR-SEM 0.100 124.8 -89.8
0.120 124.8 -89.8
1 124.5 -89.8
10 124.1 -80.0
100 123.8 -89.2
FAIR-IC 0.100 228
0.120 227
1 226
10 225.8
100 223.3
BLM-IC 0.100 428.5
0.120 428.5
1 428.7
10 427.9
100 426.8
BLM-PC 1 8

S11 measurements

S11 FAIR-SEM