User Tools

Site Tools


hi:hi:blmservicerecords

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Last revision Both sides next revision
hi:hi:blmservicerecords [2019/09/19 16:55]
p.boutachkov
hi:hi:blmservicerecords [2019/09/25 15:23]
p.boutachkov
Line 1: Line 1:
 +====== Calibration ======
 +The BLM calibration is performed with a <sup>137</sup>Cs source.
 +The source emits a single gamma at 661.7 keV. The source activity was 323.2 kBq as of 11.01.2019. Its serial number is AB-9020, 370 kBq when purchased.
 +
 +I have used 10 s integration time and set all BLM to count 30 k. (3 kHz)
 +
 +Note on the light yield. Reference Glenn Knoll 3rd edition p.235 and p226 NE102A 10 k photons per MeV. BC400 light output 0.65 of NE102A (anthrscene). Hence one can expect: 6500 photons / keV
 +
 +Bellow is a spectrum taken with the HDO6104A scope when for a BLM set to count 3 kHz (in this case XP2972 34391)
 +
 +{{:hi:hi:sn34391-137cs-1275v--00000.jpg?700|}}
 +
 +====== XP2972 PMT History ======
 +CBS - Cathode Blue sensitivity
 +ABS - For a anode blue sensitivity
 +Id - Dark current
 +
 +^ SN ^ CBS [uA/lmF] ^ ABS [A/lmF] ^ HV [V] ^ Id[nA] ^ Gain [10<sup>6</sup>] ^ Comment ^
 +| 34391 | 12.6 | 10 | 1188 | 4.98 | 0.79 | new, will be installed in TE1 |
 +| 29181 |  |  | |  |  | used, low gain, was in S07DL1R |
 +
 ====== GS06DL3U ====== ====== GS06DL3U ======
 Reason for the service: The detector counts more than 10 times compare to the detectors above, left and right of it. Reason for the service: The detector counts more than 10 times compare to the detectors above, left and right of it.
Line 19: Line 40:
     * No yellowing, signs of mechanical damage on the surface.     * No yellowing, signs of mechanical damage on the surface.
     * Under UV, no spots     * Under UV, no spots
-The Standard operating voyage is 1100 V, I<sub>l</sub> = 739 uA, counting rate bellow 5 Hz +The Standard operating voltage is 1100 V, I<sub>l</sub> = 739 uA, counting rate bellow 5 Hz 
hi/hi/blmservicerecords.txt · Last modified: 2020/12/11 13:33 by p.boutachkov